積分球LED光電色檢測操作規程

Operation procedure for the detection of the photoelectric color of the integral sphere LED

 

第一次使用儀器

The first use of the instrument

 

1)把軟件打開(kāi), 保證0.3米積分球和1.5米積分球是關(guān)閉狀態(tài),積分球內不能放任何光源。

1) to open the software, to ensure that the 0.3 meters and 1.5 meters of integrated ball is off the ball is closed, integral ball can not put any light source.

 

2)點(diǎn)擊軟件上面01.jpg清零 ok 等待此任務(wù)即可

2) click on the software to clear the OK waiting for this task

 

開(kāi)啟檢測設備

Opening test equipment

 

開(kāi)啟配電機柜總電源,開(kāi)關(guān)位置為:機柜中間面板的左側船型開(kāi)關(guān);02.jpg

Turn on the power switch of the power distribution cabinet and the switch position is: the left side of the middle panel of the cabinet;

 

開(kāi)啟機柜內部各個(gè)儀表的電源(功率計、穩壓變頻電源) ;

Power supply (power meter, voltage regulated power supply) for various instruments in the cabinet;

 

開(kāi)啟直流電源;

Open DC power supply;

 


1.5米積分球的校準測量步驟

1.5 meter measuring steps

運行桌面1.5米球的測試軟件(確認桌面0.3米球的軟件已經(jīng)關(guān)閉);03.jpg

Test software running 1.5 meters of the table, the software has been closed down for 0.3 meters;

 

04.jpg05.jpg  

1. 5米球內裝入50W標準光源;

50W standard light source in the 1.5 meter ball;

 

切換機柜右側的船型開(kāi)關(guān)到1.5米積分球;

Switch on the right side of the switch to the 1.5 meter integral ball ;

06.jpg

 

保持機柜中間面板的中間轉換開(kāi)關(guān)到“DC/直流”;

To maintain the intermediate panel of the intermediate switch to the "DC/ dc";


在軟件上點(diǎn)07.jpg進(jìn)入界面把電壓 電流 流明  色溫 設置好,數據已標準燈為準。

On the software point of entry interface, the voltage and current lumen color temperature is set up, and the data has been standard lamp.

08.jpg


數據設置好后,點(diǎn)亮標準光源09.jpg(標準燈需要預熱5分鐘時(shí)間)

After the data is set up, light up the standard light source (the standard lamp needs to warm for 5 minutes).10.jpg

標準光源點(diǎn)亮后,直接點(diǎn)校準,預熱時(shí)間過(guò)后會(huì )出現一個(gè)山峰曲線(xiàn)圖譜后,點(diǎn)擊保存退出

After the standard light source is illuminated, it will be calibrated directly. After the preheating time, there will be a map of the peak curve. Click Save and exit 11.jpg

等待校準完成后再點(diǎn)擊“開(kāi)始測試”;查看對比結果。12.jpg

Click the "Instrument calibration" button, select "spectral calibration", select "color temperature flux calibration"; enter the standard color temperature, standard flux; click on the "Calibration" and then wait for the calibration is complete click on "Start Test";. For a comparison of the results.

 

1.5米積分球校準定標完成。

1.5 meters integral sphere calibration is completed.

 

等待2分鐘,確認標準光源已經(jīng)冷卻,取下標準光源,放入燈盒中;開(kāi)始測試貴公司的產(chǎn)品

Wait for 2 minutes, confirm the standard light source has cooled, take the standard light source, into the lamp box;

 

如果產(chǎn)品是交流的話(huà)關(guān)閉穩壓直流,開(kāi)啟變頻測試電源。

Close the regulator DC, open the power supply of variable frequency test.

 

轉換機柜中間面板的中間轉換開(kāi)關(guān)到“AC/交流” 13.jpg

Conversion switch to "AC/""

 

點(diǎn)亮被測燈具,點(diǎn)擊“開(kāi)始測試” 14.jpg

Light up the lamp, click "start testing"


 

0.3米積分球的校準測量步驟

0.3 meter measuring steps

 

運行桌面0.3米球的測試軟件(確認桌面1.5米球的軟件已經(jīng)關(guān)閉);

Test software running 0.3 meters of the table, the software has been closed down for 1.5 meters;

 

0.3米球內裝入10W標準光源;15.jpg小球軟件

Load 10W standard light source within 0.3 m ball;

 

切換機柜右側的船型開(kāi)關(guān)到0.3米積分球16.jpg

Switch on the right side of the switch to the 0.3 meter integral ball

 

保持機柜中間面板的中間轉換開(kāi)關(guān)到“DC/直流” 17.jpg

To maintain the intermediate panel of the intermediate switch to the "DC/ dc"

 

點(diǎn)亮標準光源

Light standard light source

 

標準光源點(diǎn)亮后,穩定5分鐘; 軟件操作跟大球一樣。

Standard light source, stable for 2 minutes;

 

點(diǎn)擊開(kāi)始測試,查看對比結果。

Click to start the test to see the contrast results.

 

點(diǎn)擊“儀器校準”按鈕,選擇“光譜校準”,選擇“色溫光通量校準”;輸入標準色溫,標準光通量;點(diǎn)擊“校準”。等待校準完成后再點(diǎn)擊“開(kāi)始測試”;查看對比結果。

Click the "Instrument calibration" button, select "spectral calibration", select "color temperature flux calibration"; enter the standard color temperature, standard flux; click on the "Calibration" and then wait for the calibration is complete click on "Start Test";. For a comparison of the results.

 

0.3米積分球校準定標完成。

0.3 meters integral sphere calibration is completed.

 

等待2分鐘,確認標準光源已經(jīng)冷卻,取下標準光源,放入燈盒中;

Wait for 2 minutes, confirm the standard light source has cooled, take the standard light source, into the lamp box;

 

將被測LED放入相應的專(zhuān)用LED夾具,然后放入到0.3米積分球側面開(kāi)口處,固定好,夾具有(鹵鎢燈,大功率1W,3528,5050,3014等);

Will be measured LED into the corresponding special LED fixture, and then put into the 0.3 meters of the integral ball side openings, fixed, with (halogen tungsten lamp, high power 1W, 352850503014, etc.);

 

調節直流恒流電源,點(diǎn)亮被測LED;

Adjust the DC constant current power supply, light is measured LED;

 

點(diǎn)擊“開(kāi)始”按鈕,等待測量完成(光譜及電參數會(huì )自動(dòng)完成測試);

Click "start" button and wait for the measurement to finish (the spectrum and the electrical parameters will be automatically finished testing);

 

點(diǎn)擊“保存”按鈕,記錄下原始數據文件;

Click the Save button to record the raw data file;

 

導出PDF測試報告,過(guò)程:點(diǎn)擊“文件”—“導出”—“PDF”;

Export PDF test report, process: click the "file" - "export" - "PDF";

 

重復13)到19)測量下一個(gè)被測燈;

Repeat 13) to measure the next measured light by 19;

 

 18.jpg